Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/21/2010 | 1330 | 10/01/2010 | 2000 | Incorrect |
Subject: | SGP/EBBR/E13 - Net Radiation Too Large |
DataStreams: | sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | The EBBR Net Radiation (Rn) values maintained a ~200 Wm^-2 reading because of poor cable lead connections. Sensible and latent heat fluxes were therefore overestimated. Sensible (H) and latent heat (LE) fluxes can be estimated in the following way: H (new) = H/[1 + (EBBR Rn - SIRS net radiation)/SIRS net radiation] LE (new) = LE/[1 + (EBBR Rn - SIRS net radiation)/SIRS net radiation] |
Suggestions: | |
Measurements: | sgp15ebbrE13.b1: more
sgp5ebbrE13.b1: more sgp30ebbrE13.b1: more |