Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/09/2010 | 1030 | 08/04/2010 | 0200 | Suspect |
Subject: | SGP/EBBR/E20 - Incorrect Sensible and Latent Heat Fluxes - a result of low battery |
DataStreams: | sgp5ebbrE20.b1, sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | Supply voltage sometimes dropped too low at night, causing incorrect measurements of sensible and latent heat fluxes, pressure, AEM home signal, and soil heat flow. |
Suggestions: | |
Measurements: | sgp30ebbrE20.b1: more
sgp15ebbrE20.b1: more sgp5ebbrE20.b1: more |