04/17/2008 | 1400 | 04/23/2008 | 1445 | Incorrect |
Subject: | SGP/MMCR/C1 - Polarization Mode Problem |
DataStreams: | sgpmmcrspecmomC1.a0, sgpmmcrmomC1.b1 |
Description: | The cross-polarization and co-polarization modes (6 and 5) were measuring cross-polarization only. This means that one of the control signals was not hooked up correctly. The signals were rerouted on 4/23/2008. The General Mode (mode 3) is a co-polarization channel and worked during this time period. The reason why we don't use the general mode all of the time is that the co-pol and cross-pol modes are temporally closer and make a better estimation of linear depolarization ratio. |
Suggestions: | |
Measurements: | sgpmmcrmomC1.b1: more
sgpmmcrspecmomC1.a0: more |