Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/29/2008 | 0000 | 03/12/2008 | 0000 | Suspect |
Subject: | NSA/MWRP/C1 - Hardware failure- Intermittent data |
DataStreams: | nsamwrpC1.b1 |
Description: | Data are intermittent during this time due to a failure of the analog board. The calibration of V-band data may have slightly drifted due to power outages. |
Suggestions: | When the data are available their quality should be good. After the instrument is repaired the V-band data will be analyzed and may be reprocessed if a slight change in calibration is observed. |
Measurements: | nsamwrpC1.b1: more
|