Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/30/2008 | 0000 | 02/26/2008 | 2150 | Incorrect |
Subject: | SGP/SIRS/E18 - NIP condensation problem |
DataStreams: | sgpsirs20sE18.a0, sgpsirsE18.b1 |
Description: | NIP at E18 was measuring impossible values due to condensation on inside of NIP window. PM replaced NIP 2/26/2008 2150. Measured data are not recoverable. Use component sum method to reconstruct NIP data. NIP = (Global-Diffuse)/cos(ZA) |
Suggestions: | |
Measurements: | sgpsirs20sE18.a0: more
sgpsirsE18.b1: more |