Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/20/2007 | 1849 | 03/05/2008 | 0200 | Incorrect |
Subject: | TWP/TSI/C2 - Mirror continuously rotating |
DataStreams: | twptsiskyimageC2.a1, twptsicldmaskC2.a1, twptsiskycoverC2.b1, twptsimovieC2.a1 |
Description: | The mirror began to continuously spin. Diagnostics were performed on the encoder, control board, and Ethernet adaptor. The instrument was removed on 1/2/2008 and shipped to Darwin for repair. The encoder and control board were replaced and the instrument was reinstalled. |
Suggestions: | |
Measurements: | twptsiskycoverC2.b1: more
twptsimovieC2.a1: more twptsiskyimageC2.a1: more twptsicldmaskC2.a1: more |