Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/12/2005 | 1530 | 03/28/2008 | 1722 | Suspect |
Subject: | SGP/IRT/E15 - Instrument noise problem |
DataStreams: | sgpirtE15.b1, sgpirtE15.00, sgpirt200msE15.a1 |
Description: | The data was much noisier than other nearby sites (a factor of 4-5 times more). The connections and communication and other settings were checked. The IRT was replaced on 1/29/2008 at 17:37 but the response time was incorrectly set to 0.01 s. This problem was corrected when the setting was changed to 0.10 s. |
Suggestions: | |
Measurements: | sgpirtE15.00: more
sgpirtE15.b1: more sgpirt200msE15.a1: more |