Subject: | SGP/TWR/C1 - Southeast Side (10X) Temperature and RH Spikes |
DataStreams: | sgp30twr10xC1.b1, sgp1twr10xC1.b1
|
Description: | Southeast side 25 m and 60 m (10X datastreams) carriage connectors may not have been well
connected; temperatures and relative humidities spiked sometimes. This problem occurred
after the carriages were lowered on 12/04. |
Suggestions: | Use the data from the West side of the tower. |
Measurements: | sgp30twr10xC1.b1: - sd_vap_pres_25m
- temp_25m
- vap_pres_25m
- sd_rh_25m
- sd_temp_60m
- sd_vap_pres_60m
- vap_pres_60m
- sd_rh_60m
- sd_temp_25m
- rh_60m
- temp_60m
- rh_25m
more
sgp1twr10xC1.b1: - vap_pres_60m
- vap_pres_25m
- temp_25m
- rh_25m
- temp_60m
- rh_60m
more
|