Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/27/2007 | 2000 | 08/05/2007 | 1300 | Suspect |
Subject: | SGP/MFRSR/E5 - Heater problem |
DataStreams: | sgpmfrsrE5.a0, sgpmfrsrlangleyE5.c1, sgpmfrsrE5.b1, sgpmfrsrlangplotE5.c1 |
Description: | This site had a new instrument and logging system installed May 07. The first data we retrieved were in June and the instrument heater was not functioning. Maintenance on 8/5/2007 resolved the issue. The head is designed to operate at 40 deg C, and we like to see it stay within a few degrees of 40 deg C during normal operation. While the irradiance data look OK, there could be subtle issues resulting from the head temperature not being stable. |