Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/03/2007 | 2005 | 01/23/2007 | 1930 | Incorrect |
Subject: | SGP/SIRS/C1 - Upwelling LW cable problems |
DataStreams: | sgpsirsC1.00, sgpsirsC1.b1, sgpsirs20sC1.a0 |
Description: | Troubleshooting of the upwelling shortwave instrument on Jan 3rd 2007 caused problems with the co-located upwelling longwave hemispheric irradiance Eppley PIR instrument (aka UIR). Data dropouts persisted from this date to the final resolution of both shortwave and longwave instrument problems. Rodents had chewed through the instrument cables. The cables were replaced 01/23/2007 1930GMT. |
Suggestions: | Data is not recoverable. disregard this data. |
Measurements: | sgpsirsC1.b1: more
sgpsirsC1.00: more sgpsirs20sC1.a0: more |