Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/30/2006 | 1500 | 11/21/2006 | 1800 | Suspect |
Subject: | SGP/MFRSR/E16 - Heater problem |
DataStreams: | sgpmfrsrE16.a0, sgpmfrsrlangleyE16.c1, sgpmfrsrE16.b1, sgpmfrsrlangplotE16.c1 |
Description: | The heater was not functioning during this period. MFRSRs are designed to operate at 40 deg C. Since the heater was not functional during this period all irradiance data are suspect. |
Suggestions: | Use data with caution |
Measurements: | sgpmfrsrlangleyE16.c1:
sgpmfrsrlangplotE16.c1:
sgpmfrsrE16.a0:
sgpmfrsrE16.b1:
|