Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/01/2005 | 0000 | 04/27/2007 | 1618 | Incorrect |
Subject: | SGP/IRT/E19 - Reprocess: Calibration error |
DataStreams: | sgpirtE19.b1, sgpirt200msE19.a1 |
Description: | The IRT was measuring ~20K higher sky temperatures than at other nearby extended facilies. The calibration of the instrument was incorrect. The problem was corrected by replacing the IRT. The offset in the calibration will be determined and the data will be corrected. |
Suggestions: | |
Measurements: | sgpirt200msE19.a1: more
sgpirtE19.b1: more |