Description: | At the beginning of November, the K-band channel calibration started to drift. The MWRP
brightness temperatures were about 1.5 K higher than the MWR-measured brightness
temperatures. By the half of the month the differences were more than 2 K to become about 4 K by
the end of November. On Nov 27 new calibration coefficients were derived and uploaded.
Data for 20061101-20061127 have been reprocessed with the revised calibrations and are