Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/27/2006 | 0608 | 11/27/2006 | 2226 | Incorrect |
Subject: | NSA/MET/TWR/C2 - 2m Temp/RH sensor blocked with snow |
DataStreams: | nsametC2.b1, nsatwrC2.b1 |
Description: | The 2m RH, Dew Point and Vapor Pressure values are incorrect due to the intake of the aspirated shield being blocked with snow and ice. The RH% values are artificially larger than the 5m and Chilled Mirror values. |
Suggestions: | Use the 5m or chilled mirror data in place of the 2m data. |
Measurements: | nsametC2.b1: more
nsatwrC2.b1: more |