Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/20/2004 | 2200 | 12/21/2005 | 2136 | Does not affect quality |
Subject: | TWP/SKYRAD/C3 - Reprocessed: Longwave Calibration error |
DataStreams: | twpskyrad60sC3.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place were reprocessed to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Suggestions: | Reprocessed data were archived 20100211. |
Measurements: | twpskyrad60sC3.b1: more
|