Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/22/2006 | 1300 | 05/30/2006 | 2020 | Incorrect |
Subject: | SGP/SIRS/E8 - Instrument cable problem |
DataStreams: | sgpsirsE8.b1, sgpsirs20sE8.a0 |
Description: | The short_direct_normal, down_short_diffuse_hemisp and up_short_hemisp data are missing or are spiking to negative numbers. On 20060530, technicians discovered cows had chewed on the instrument cables. |
Suggestions: | Disregard this data |
Measurements: | sgpsirsE8.b1: more