Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/16/2006 | 1705 | 04/13/2006 | 1715 | Does not affect quality |
Subject: | SGP/SIRS/E24 - Reprocessing Notice (RID 257) |
DataStreams: | sgpsirs20sE24.a0, sgpsirsE24.b1, sgpsirsE24.00 |
Description: | See below for details of reprocessing that was completed on 20160127. Original DQR category: 4 (Red, Incorrect). ----------------------------------------------------------- During the PIR coefficient change from 4/coeffs to 1/coeff, the downwelling diffuse hemispheric instrument (Eppley 8-48) coefficient was inadvertently changed to 122.85 and should have been 112.85 W/m2/mV. Perform the following calculation to correct the data, DD=downwelling diffuse instrument: DD(new)[W/m2]=( DD(old)[W/m2]/122.85[W/m2\mV] )* 112.85[W/m2/mV] |
Suggestions: | |
Measurements: | sgpsirsE24.b1: more
sgpsirsE24.00: more sgpsirs20sE24.a0: more |