Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/01/2005 | 0100 | 11/04/2005 | 0600 | Incorrect |
Subject: | TWP/MMCR/C3 - MMCR Power Monitor failure |
DataStreams: | twpmmcrpowC3.a1, twpmmcrcalC3.a1 |
Description: | During this period, the transmit power monitor on the TWTA was not working. This resulted in the power level reading staying at a steady state (approximately 49 dBm). Normal transmitter fluctuations are expected to be on the order of +/- 0.3 dBm. |
Suggestions: | Reflectivity values during this period have a +/- 0.3 dBZ uncertainty. |
Measurements: | twpmmcrcalC3.a1: more
twpmmcrpowC3.a1: more |