Subject: | NSA/MWR/C1 - Instrument problem, Heater problem |
DataStreams: | nsamwrlosC1.b1, nsamwrC1.00
|
Description: | The heater of the MWR was ON all the time even when the moisture sensor was dry. Jim
Liljegren suspected a problem with the relay that activates the heater. The suggested action
was to send the faulty assembly to SGP for repair and get a spare unit from SGP.
The spare unit was sent on 9/15 and was installed on 10/18. |
Suggestions: | Data should not be affected by this problem |