Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/20/2005 | 1800 | 08/05/2005 | 1935 | Suspect |
Subject: | SGP/MWR/B1 - thermal instability |
DataStreams: | sgpmwrlosB1.b1, sgpmwrtipB1.a1 |
Description: | During periods of high ambient temperature, the instrument may have been thermally unstable. The problem was corrected by increasing the RF deck temperature from 50 to 52 C. |
Suggestions: | The data are suspect when the blackbody temperature, tkbb, exceeds 318 K. |
Measurements: | sgpmwrtipB1.a1: more