Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/09/2005 | 0849 | 01/18/2005 | 1900 | Incorrect |
Subject: | SGP/SIRS/E15 - Failed NIP cable |
DataStreams: | sgpsirsE15.00, sgpsirs20sE15.a0, sgpsirsE15.b1 |
Description: | After the ice storm, the SHORT_DIRECT_NORMAL values were bad or missing continuously. Corrective Maintenance on 01/18 found the NIP cable had broken inside the Bendix connector. The NIP cable was repaired and returned to service. |
Suggestions: | Data is not recoverable. Disregard this data. |
Measurements: | sgpsirsE15.00: more
sgpsirs20sE15.a0: more sgpsirsE15.b1: more |