Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/30/2003 | 1545 | 11/13/2003 | 1545 | Incorrect |
Subject: | SGP/EBBR/E26 - AEM Home Signals Incorrect, Surface Fluxes Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | During a preventative maintenance visit to perform the 6-month calibrations, the home15 and home30 outputs began reading the same as one another. A fuse replacement and lubrication of the AEM sprockets on 11/13 corrected the problem. During the interval specified, the home15, home30, and sensible and latent heat fluxes are incorrect. |
Suggestions: | There is no work around to this problem. |
Measurements: | sgp5ebbrE26.b1: more
sgp15ebbrE26.b1: more sgp30ebbrE26.b1: more |