Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/03/2003 | 0000 | 07/28/2003 | 2359 | Suspect |
Subject: | SGP/MFRSR/E1 - Instrument Noise Problem |
DataStreams: | sgpmfrsrE1.b1, sgpmfrsrE1.a0 |
Description: | Intermittent spikes are evident in the data for this time period. These spikes are due to saturation of the instrument signal in both the diffuse and total (global) measurements as a result of data logger problems. The corresponding direct data can be observed to have an erroneous sudden drop in signal when the spikes in the diffuse and total measurement are observed. |
Suggestions: | Screened data before use |
Measurements: | sgpmfrsrE1.b1: more
sgpmfrsrE1.a0: more |