Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/19/2003 | 2324 | 07/21/2003 | 0824 | Suspect |
Subject: | NSA/MWR/C2 - Thermally unstable |
DataStreams: | nsamwrlosC2.b1 |
Description: | The MWR became thermally unstable during periods when the ambient temperature exceeded about 28C. This situation was corrected when the mixer temperature was manually increased to about 35C. Data are suspect when the blackbody temperature, tkbb, exceeds 296 K. |
Suggestions: | |
Measurements: | nsamwrlosC2.b1: more
|