Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/24/2003 | 1100 | 07/09/2003 | 1802 | Incorrect |
Subject: | SGP/MFRSR/E20 - Failed Sensor results in erratic data |
DataStreams: | sgpmfrsrlangplotE20.c1, sgpmfrsrE20.a0, sgpmfrsrlangleyE20.c1, sgpmfrsrE20.00, sgpmfrsrE20.b1 |
Description: | The 415 nm data was suspiciously erratic, noisy, and/or high. The other channels also seem somewhat suspicious, especially on 06/27 and 06/28 when readings were around 0 on all filters for a lengthy period of time (which did not agree with surrounding facilities). |
Suggestions: | Do not use data |
Measurements: | sgpmfrsrlangplotE20.c1:
sgpmfrsrE20.00:
sgpmfrsrlangleyE20.c1:
sgpmfrsrE20.a0:
sgpmfrsrE20.b1:
|