Description: | > From an email message from John Goldsmith (SNL)
The site personnel had some alignment problems restarting the Raman
lidar after they stopped it yesterday (4/17) morning to clean the window.
They finally got it running late in the afternoon, but they left a
terminated cable hanging from a second output on the Aerosol High amplifier
(used with a scope to find the beam when the alignment is way off).
Electronics can be picky; that cable screwed up the Aerosol High data,
filling the column with large numbers (~60,000); the other six columns are
fine. The cable was removed at my request this morning; the first dataset
with good Aerosol High values is 71080842.DAT, #992 in the sequence, start
time 14:36:17 Z. |