Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/25/1996 | 0000 | 08/02/2000 | 1450 | Incorrect |
Subject: | SGP/SWATS/E1 - Reprocess: Wrong Calibration |
DataStreams: | sgpswatsE1.a1, sgpswatsE1.b1, sgpswatsE1.00 |
Description: | Two sensor connections were switched. The East 125-cm sensor was plugged into the East 85-cm sensor's connection, and vice versa. Technicians corrected the problem on 1440 GMT, 08/02/00, and returned the instrument to service 10 minutes later at 1450 GMT. As the datalogger was confusing the 2 sensors, incorrect calibrations were applied. The correct calibrations can be applied to the raw data to obtain the correct temperature rises, soil moisture, etc. However, this is a rather large amount of work to apply toward just 1 measurement profile. This has been identified as a medium priority reprocessing task. |