Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/12/2002 | 0738 | 06/25/2002 | 2359 | Does not affect quality |
Subject: | TWP/SKYRAD/C3 - Reprocess: PIR identified as unshaded is really shaded |
DataStreams: | twpskyrad60sC3.a1 |
Description: | From the initial installation of the Dqrwin SKYRAD suite if instruments, two shaded PIRs were installed rather than one shaded and one unshaded PIR as documented in the netcdf header information. See P011109.2 for more discussion of this issue. |
Suggestions: | Disregard the data description in the netcf header |
Measurements: | twpskyrad60sC3.a1: more
|