Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/25/2001 | 1545 | 12/28/2001 | 1931 | Suspect |
Subject: | SGP/EBBR/E13 - Left T/RH Probe Temperature Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | During much of the indicated period the Left T/RH Temperature was incorrect. Because the home signal was between 0 and 3, the temperature was incorrect when the AEM was at the bottom position, but was correct when the AEM was at the top position. This was caused by a short created when the sensor was mispositioned in the aspirator housing and enough strain on the cable for it to bring it into contact with the housing in the bottom position. |
Suggestions: | |
Measurements: | sgp15ebbrE13.b1: more
sgp5ebbrE13.b1: more sgp30ebbrE13.b1: more |