Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/21/2000 | 1506 | 09/21/2000 | 1517 | Incorrect |
02/16/2000 | 2210 | 02/16/2000 | 2250 | Incorrect |
10/02/2000 | 2052 | 10/02/2000 | 2107 | Incorrect |
09/26/2000 | 1426 | 09/26/2000 | 1434 | Incorrect |
09/21/2000 | 1423 | 09/21/2000 | 1434 | Incorrect |
10/02/2000 | 1732 | 10/02/2000 | 1752 | Incorrect |
10/03/2000 | 1406 | 10/03/2000 | 1421 | Incorrect |
09/06/2001 | 1900 | 09/06/2001 | 2000 | Incorrect |
Subject: | SGP/MWR/C1 - calibration checks |
DataStreams: | sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgpmwrlosC1.a1 |
Description: | The MWR calibration was checked by a cryogenic (liquid nitrogen) blackbody target or (in the case of the Sept 2001 time period) by pointing the mirror at an internal blackbody target. The brightness temperatures and vap and liq retrievals are not representative of the sky/atmosphere. |
Suggestions: | Do not use data. |
Measurements: | sgpmwrlosC1.a1: more
sgpmwrtipC1.a1: more sgpmwrlosC1.b1: more |