Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/19/1998 | 0000 | 08/22/1998 | 0000 | Suspect |
Subject: | TWP/CEIL/C1 - Ceilometer clock drift |
DataStreams: | twpceilC1.b1, twpceilC1.00 |
Description: | During this period the ceilometer was using a software version that provided time stamps that were based on the ceilometer clock and not the computer clock that was continually sychronized with GPS. This lead to times that were erroneous by as much as 2 minutes. For most applications 2 minutes is an acceptable error, but for some applications this may be a problem. |
Suggestions: | User beware. |
Measurements: | twpceilC1.b1: more
twpceilC1.00: more |