Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/24/2016 | 1830 | 03/29/2016 | 1920 | Incorrect |
Subject: | SGP/AOS/C1 - No flow control through impactor |
DataStreams: | sgpaospsap3w1mC1.b1, sgpaosnephwetC1.a1, sgpaosnephdryC1.b1, sgpaosnephdryC1.a1, sgpaospsap3wC1.a1, sgpaosclap3wC1.a1, sgpaosnephwet1mC1.b1, sgpaosnephdry1mC1.b1, sgpaosclap3wC1.b1, sgpaosnephwetC1.b1, sgpaospsap3wC1.b1 |
Description: | New control board installed and system not rebooted to initiate communications. As a result there was no flow control through the impactors. This affects all instruments downstream of the impactors. |
Suggestions: | Do not use data |
Measurements: | sgpaosnephdry1mC1.b1: more
sgpaosnephwet1mC1.b1: more sgpaospsap3w1mC1.b1: more sgpaosclap3wC1.b1: more sgpaosclap3wC1.a1: more sgpaosnephwetC1.b1: more sgpaospsap3wC1.a1: more sgpaosnephwetC1.a1: more sgpaospsap3wC1.b1: more sgpaosnephdryC1.a1: more sgpaosnephdryC1.b1: more |