Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/30/2013 | 1605 | 06/19/2014 | 2000 | Does not affect quality |
Subject: | MAO/AETH/NEPHELOMETER/PASS/PSAP/HTDMA/NEPHELOMETER/PASS/S1 - Impactor settings reversed |
DataStreams: | maoaosnephdryS1.a1, maoaosaethS1.a1, maoaospsap3wS1.b1, maoaospsap3wS1.a1, maoaosnephwetS1.b1, maoaosnephdryS1.b1, maoaospsap3wS1.01, maoaospass3wS1.a1, maoaosaethS1.00 |
Description: | The impactor was reporting the opposite state of what was being measured. I.e. measuring 10um but reporting 1um. |
Suggestions: | The correct state can be determined by reviewing the Nephelometer pressure (10um=higher pressure, 1um=lower pressure) |
Measurements: | maoaospass3wS1.a1: more
maoaospsap3wS1.a1: more maoaospsap3wS1.01: more maoaospsap3wS1.b1: more maoaosaethS1.a1: more maoaosaethS1.00: more maoaosnephdryS1.b1: more maoaosnephdryS1.a1: more maoaosnephwetS1.b1: more |