Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/12/2014 | 1845 | 11/14/2014 | 1356 | Suspect |
Subject: | MAO/IMPACTOR/S1 - Impactor not switching |
DataStreams: | maoaosimpactorS1.a1 |
Description: | The impactor stopped in the 1 micron size cut, and the data in this time range is valid for a 1 micron size. There is no 10 micron size cut data since the impactor had stopped. |
Suggestions: | Use data as 1 micron size cut data only. |
Measurements: | maoaosimpactorS1.a1: more
|