Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2014 | 0000 | 11/05/2014 | 0641 | Incorrect |
Subject: | TWP/DL/C3 - Fault in RF attenuator |
DataStreams: | twpdlC3.00, twpdlppiC3.b1, twpdlppiC3.a0, twpdlfptC3.a0, twpdlcal1C3.b1, twpdlcal1C3.a0, twpdlcal2C3.b1, twpdlfptC3.b1, twpdlcal2C3.a0 |
Description: | The instrument was shipped to the vendor (Halo Photonics) for repair. The root cause of the problem was traced to an intermittent fault in an RF attenuator in the transmitter. The RF attenuator was replaced. In addition to the repair, several upgrades were performed. These included the installation of a new PC with Win7, new control software, and the installation of a tip-tilt sensor. |
Suggestions: | Do not use data |
Measurements: | twpdlcal2C3.a0: more
twpdlcal2C3.b1: more twpdlfptC3.a0: more twpdlC3.00: more twpdlfptC3.b1: more twpdlcal1C3.a0: more twpdlcal1C3.b1: more twpdlppiC3.b1: more twpdlppiC3.a0: more |