Subject: | SGP/SIRS/E15 - Broken wire resulting in missing data |
DataStreams: | sgpsirsE15.b1, sgpsirs20sE15.a0
|
Description: | The Short Direct Normal (NIP) instrument had a broken wire in the connector. The broken
wire was repaired on 7/18/14. |
Suggestions: | The short_direct_normal data can be approximated from the two other shortwave
datastreams available using the following relationship:
short_direct_normal = (down_short_hemisp -down_short_diffuse)/cos(zenith angle) |
Measurements: | sgpsirs20sE15.a0: more
sgpsirsE15.b1: - short_direct_normal
- short_direct_normal_max
- short_direct_normal_min
- short_direct_normal_std
more
|