Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/30/2013 | 1605 | 06/19/2014 | 2000 | Incorrect |
Subject: | MAO/AOS/S1 - Impactor Setting Reversed |
DataStreams: | maoaosimpactorS1.a1 |
Description: | The impactor was reporting the opposite state of what was being measured. I.e. measuring 10um but reporting 1um. |
Suggestions: | The correct state can be determined by reviewing the Nephelometer pressure (10um=higher pressure, 1um=lower pressure) |
Measurements: | maoaosimpactorS1.a1: more
|