Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/01/2014 | 0615 | 03/17/2014 | 2057 | Incorrect |
Subject: | TWP/SKYRAD/C1 - IRT Configuration |
DataStreams: | twpskyrad60sC1.b1 |
Description: | The IRT was upgraded and the scale factors for both the IRT and datalogger program were changed accordingly but the startup parameters for the serial collection script, which automatically reconfigure the analog output of the IRT, were not changed. The parameter in the startup file was corrected. |
Suggestions: | The sky_ir_temp measurements can be corrected by applying: (25/26)(sky_ir_temp-173.2K) + 213.2K |
Measurements: | twpskyrad60sC1.b1: more
|