Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/18/2011 | 0000 | 02/07/2014 | 2359 | Incorrect |
Subject: | NSA/KAZR/C1 - Reprocess: Incorrect calibration coefficients |
DataStreams: | nsakazrspeccmaskgecopolC1.a0, nsakazrspeccmaskhicopolC1.a0, nsakazrmdC1.a1, nsakazrspeccmaskmdxpolC1.a0, nsakazrspeccmaskmdcopolC1.a0, nsakazrspeccmaskhixpolC1.a0, nsakazrhiC1.a1, nsakazrspeccmaskgexpolC1.a0, nsakazrgeC1.a1, nsakazrauxC1.a0 |
Description: | reflectivity between long and short pulse did not agree with each other after analyzing data. Data will be reprocessed. |
Suggestions: | Do not use data until data has been reprocessed. |
Measurements: | nsakazrgeC1.a1: more
nsakazrauxC1.a0: more nsakazrspeccmaskmdcopolC1.a0: more nsakazrhiC1.a1: more nsakazrspeccmaskhicopolC1.a0: more nsakazrspeccmaskhixpolC1.a0: more nsakazrmdC1.a1: more nsakazrspeccmaskmdxpolC1.a0: more nsakazrspeccmaskgexpolC1.a0: more nsakazrspeccmaskgecopolC1.a0: more |