Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/25/2011 | 0000 | 08/03/2011 | 0000 | Suspect |
Subject: | SGP/AOS/C1 - Nephelometer shutter error |
DataStreams: | sgpnoaaaosfitrhC1.b1, sgpaipavg1ogrenC1.c1, sgpnoaaaosC1.b1, sgpaip1ogrenC1.c1, sgpnoaaaosavgC1.b1, sgpaosnephdryC1.a1 |
Description: | The neph backscatter shutter started to fail which resulted in large background values in the dry nephelometer. The data is poor quality even though it may look ok. Instead use the data in the wet nephelometer. Note that during some of this time period the RH value inside of this neph ramped up and down so use data at the low RH values. |
Suggestions: | Use wet neph data as described |
Measurements: | sgpnoaaaosfitrhC1.b1: more
sgpnoaaaosavgC1.b1: more sgpaosnephdryC1.a1: more sgpaip1ogrenC1.c1: more sgpnoaaaosC1.b1: more sgpaipavg1ogrenC1.c1: more |