Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/12/2011 | 0000 | 06/14/2011 | 1750 | Incorrect |
Subject: | SGP/BRS/C1 - SW Direct Normal incorrect |
DataStreams: | sgpbrsC1.b1, sgpbrs20sC1.a0 |
Description: | Apparent lightning strike in the instrument cluster area resulted in a failed NIP. Data are missing or incorrect. The SIRS was due for calibration change outs so the US, DS, DD and NIP were replaced. The updated program was loaded at approximately 06/14/2011 1739 GMT. An estimate of direct normal values can be derived using the following formula: short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle) |
Suggestions: | |
Measurements: | sgpbrs20sC1.a0: more
sgpbrsC1.b1: more |