Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/11/2011 | 0000 | 01/27/2011 | 1920 | Suspect |
Subject: | SBS/MWR3C/M1 - Intermittent thermal instability |
DataStreams: | sbsmwr3cM1.b1 |
Description: | The radiometers reset the thermal control every time the software was started at 00:00 UTC. It took about an hour to reestablish the thermal control of the instrument after each reset. On Jan 27 a workaround was introduced to avoid this behavior. The software will be modified at the next release. |
Suggestions: | |
Measurements: | sbsmwr3cM1.b1: more
|