Subject: | SGP/SIRS/E24 - Loose connections |
DataStreams: | sgpsirs20sE24.a0, sgpsirsE24.b1
|
Description: | The down_short_hemisp, down_long_hemisp_shaded & up_long_hemisp were reporting large
negative values. A loose ground wire was reconnected and the MUX was replaced. |
Suggestions: | Disregard the specified data during this time period |
Measurements: | sgpsirsE24.b1: - up_long_hemisp_max
- up_long_hemisp_std
- up_long_netir
- down_short_hemisp_min
- down_long_hemisp_shaded_max
- down_long_hemisp_shaded_std
- up_long_hemisp
- inst_down_long_shaded_case_temp
- down_short_hemisp_max
- down_long_netir
- inst_up_long_case_temp
- down_short_hemisp_std
- inst_up_long_dome_temp
- down_long_hemisp_shaded_min
- down_long_hemisp_shaded
- up_long_hemisp_min
- down_short_hemisp
- inst_down_long_shaded_dome_temp
more
sgpsirs20sE24.a0: - inst_up_long_case_resist
- inst_global
- inst_up_long_hemisp_tp
- inst_up_long_dome_resist
- inst_up_short_hemisp
- inst_down_long_hemisp_shaded_tp
- inst_down_long_shaded_case_resist
- inst_down_long_shaded_dome_resist
more
|