Subject: | SGP/MFR10M/C1 - Artifact in data |
DataStreams: | sgpmfr10mC1.b1, sgpmfr10mC1.a0
|
Description: | A cable was hanging under the instrument and was in the field of view of the instrument.
This resulted in some erroneous increases in the data, frequently around 22GMT. These
increases are likely from specular reflections off the cable. |
Suggestions: | Use MFR25m data or screen for errors by comparing to MFR25m |
Measurements: | sgpmfr10mC1.b1: - up_hemisp_broadband
- lat
- up_hemisp_narrowband_filter6
- up_hemisp_narrowband_filter4
- up_hemisp_narrowband_filter5
- cosine_solar_zenith_angle
- alt
- logger_volt
- up_hemisp_narrowband_filter2
- up_hemisp_narrowband_filter3
- base_time
- up_hemisp_narrowband_filter1
- time_offset
- lon
- callang_flags
- head_temp
more
sgpmfr10mC1.a0: - base_time
- up_hemisp_broadband
- lon
- logger_volt
- callang_flags
- lat
- time_offset
- alt
- up_hemisp_narrowband_filter1
- up_hemisp_narrowband_filter3
- up_hemisp_narrowband_filter2
- up_hemisp_narrowband_filter5
- up_hemisp_narrowband_filter4
- up_hemisp_narrowband_filter6
- head_temp
more
|