Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/03/2003 | 0000 | 08/27/2003 | 2359 | Suspect |
Subject: | SGP/MWR/B1 - thermal instability |
DataStreams: | sgpmwrlosB1.a1, sgpmwrlosB1.b1 |
Description: | During periods of high ambient temperature, the instrument may have been thermally unstable. |
Suggestions: | The data are suspect when the blackbody temperature, tkbb, exceeds 317 K. |
Measurements: | sgpmwrlosB1.a1: more