Description: | Beginning on 06/03, the 415 nm channel appears to be erratic or noisy at times, which does
not coincide with the other channels. Maintenance on 07/02 corrected this problem.
head_temp readings occasionally spike or are erratic, especially between 1500 and 2000
UTC. On 06/17, head_temp readings became ambient due to a bad signal cable connection.
Maintenance on 07/02 ended this problem as well.
Due to the head_temp readings becoming largely out of tolerance, all channel readings
should be considered questionable, especially after 06/17. |