Subject: | TWP/GNDRAD/C1 - GNDRAD logger configured as SMET |
DataStreams: | twpgndrad60sC1.b1, twpgndrad60sC1.a1, twpgndradC1.00, twpgndrad20sC1.a1
|
Description: | the configuration for the SMET was inadvertently loaded into the GNDRAD at Manus at
03/03/2003 17:00 GMT. This error was found and corrected on 03/13/2003 at 21:30 GMT. |
Suggestions: | Because the wrong channels were addressed with incorrect sampling coefficients, it
would be very difficult (and in some cases impossible) to reconstruct the data for this
week and 3 day period. |
Measurements: | twpgndradC1.00: - Raw data stream - documentation not supported
more
twpgndrad20sC1.a1: - inst_up_long_case_resist
- lat
- alt
- pir1_dome_therm
- filename
- time_offset
- lon
- pir1_voltage
- pir1_case_therm
- inst_up_long_hemisp_tp
- base_time
- inst_up_long_dome_resist
more
twpgndrad60sC1.b1: - psp1_mean
- internal_voltage
- irt1_max
- base_time
- net
- bit_data
- lat
- psp1_sd
- net_std
- up_short_hemisp_std
- alt
- psp1_min
- up_long_hemisp
- sfc_ir_temp_min
- up_long_hemisp_std
- net_min
- irt1_min
- sfc_ir_temp_max
- time_offset
- pir1_sd
- irt1_sd
- lon
- filename
- external_voltage
- up_short_hemisp_min
- irt1_mean
- psp1_max
- net_max
- sfc_ir_temp
- up_short_hemisp_max
- pir1_min
- net_sd
- net_mean
- up_long_hemisp_max
- up_short_hemisp
- pir1_max
- logger_temp
- sfc_ir_temp_std
- pir1_mean
- up_long_hemisp_min
more
twpgndrad60sC1.a1: - up_long_hemisp_max
- irt1_sd
- pir1_min
- up_long_hemisp
- up_long_hemisp_min
- filename
- pir1_max
- sfc_ir_temp
- up_short_hemisp
- net
- psp1_mean
- lat
- sfc_ir_temp_std
- psp1_max
- net_std
- up_short_hemisp_max
- bit_data
- net_sd
- time_offset
- net_min
- up_long_hemisp_std
- net_mean
- up_short_hemisp_min
- psp1_min
- sfc_ir_temp_max
- irt1_max
- base_time
- pir1_sd
- external_voltage
- alt
- irt1_mean
- irt1_min
- pir1_mean
- net_max
- lon
- internal_voltage
- psp1_sd
- up_short_hemisp_std
- sfc_ir_temp_min
- logger_temp
more
|