Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/21/1999 | 0000 | 05/23/2001 | 0610 | Suspect |
Subject: | NSA/SKYRAD/C2 - Reprocess: IRT insufficiently heated |
DataStreams: | nsaskyrad60sC2.b1, nsaskyrad60sC2.a1 |
Description: | IRT had insufficient heating to maintain an internal reference temperature above 0 degrees C. Measurements of sky temperature were over-estimated when instrument was below freezing. Data will be reprocessed, but users can correct data using the following correction factor: If T_reference < -5°C, then IRT_corrected = (IRT_original - 32.993K)/0.87238 where T_reference can be estimated with the ambient temperature (e.g. inst_down_long_hemisp_dome_temp or inst_down_long_dome_temp) |
Suggestions: | |
Measurements: | nsaskyrad60sC2.a1: more
nsaskyrad60sC2.b1: more |