Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/17/2000 | 0101 | 11/29/2001 | 2359 | Suspect |
Subject: | SGP/NIMFR/C1 - Centrovision filter leakage |
DataStreams: | sgpnimfrC1.b1 |
Description: | Testing of the batch of centrovision filters being used in the instrument indicate that the first four filters leak 'out of band' light. The leakage for each filter is (1)415nm 3%, (2)500nm 1-2%, (3)615nm 1-2%, (4)673nm 3%. Acceptable leakage is less than 0.5%. This leakage causes an error in the irradiance values (too high) at these four wavelengths and the optical depths calculated using this data will be too high |
Suggestions: | Do not use data |
Measurements: | sgpnimfrC1.b1: more
|