Description: | Soil heat flow #2 measurements had a negative offset. Negative values were too small and
negative values occurred when positive values should have. This can occur when there is
a poor wire connection or a J-panel over-voltage protection device has failed.
Soil heat flow #4 was incorrect because it's wires were loose from the connection
terminals.
Average soil heat flow, sensible heat flux, and latent heat flux were affected;
these can be recalculated using the remaining good sets of soil sensors as follows:
ave_shf = (g1 + g3 + g5)/3
e = -(q + ave_shf)/(1 + bowen)
h = -(q + e + ave_shf) |